AgaMatrix Jazz Wireless 2 Owner's Manual page 47

Sync results bluetooth
Table of Contents

Advertisement

since you are the only person operating them.
· If the meter is being operated by a second person who
is providing testing assistance to the user, the meter and
lancing device should be cleaned and disinfected prior to
being handled by the second person.
The disinfection instructions listed below were validated for
260 cleaning and 260 disinfection cycles, for a total of 520
wipes (260 cleaning wipes plus 260 disinfecting wipes) on the
lancing device and meter.
The life of the AgaMatrix Jazz Wireless 2 Meter, defined as
5,000 test strip insertions, is between 3 to 5 years, depending
on use. The number of cleaning and disinfecting wipes used
simulates cleaning and disinfecting your meter once a week
for 5 years.
What can you use to clean and disinfect your meter and
lancing device?
See Table 1 for the suggested disinfecting wipe. You should
use wipes that are pre-saturated with disinfectant. Do not im-
merse your meter or body of the lancing device in any liquid
solution.
If you suspect your meter is not working properly after
disinfecting, verify that your system is working properly by
performing a control solution test. If your control solution
test result falls out of control range, call Customer Service.
If you notice any signs of deterioration on the meter or
lancing device, such as clouding on the meter, corrosion or
erosion of meter or lancing device plastic housing, cracking
of plastic housing, display, or button) or if the meter does not
45

Advertisement

Table of Contents
loading

Table of Contents