AgaMatrix Jazz Wireless 2 Owner's Manual page 10

Sync results bluetooth
Table of Contents

Advertisement

Meter, a sample of blood obtained by fingerstick is applied
to the test strip. The blood sample size for the meter is a
minimum of 0.5 μL.
The blood reacts with a reagent on the test strip that contains
glucose oxidase, an electron mediator, and other materials.
The chemical reaction with glucose that takes place on the
strip generates an electrical signal that is measured and pro-
cessed by the meter.
The meter computes a glucose concentration and displays
the result. The measurement range of the system is 20 to
600 mg/dL glucose. Capillary whole blood samples with a
hematocrit range of 20% to 60% can be measured.
The AgaMatrix Jazz Wireless 2 Meter is calibrated to provide
a glucose concentration for a capillary whole blood sample
that is equivalent to the plasma glucose concentration of that
sample.
AgaMatrix Jazz Test Strips
AgaMatrix Jazz Test Strips are for use with the AgaMatrix Jazz
Wireless 2 Blood Glucose Meter to quantitatively measure
glucose of fresh capillary whole blood drawn from the the
fingertip.
AgaMatrix Control Solutions
AgaMatrix Control Solutions Levels 2 and 4 are for use with the
AgaMatrix Jazz Wireless 2 Blood Glucose Meter and AgaMatrix
Jazz Test Strips to check that the meter and test strips are
8

Advertisement

Table of Contents
loading

Table of Contents