AgaMatrix Jazz Wireless 2 Owner's Manual page 31

Sync results bluetooth
Table of Contents

Advertisement

The control solution test result is stored in the meter's
memory but will not transfer to the App during syncing.
Example Result:
Compare the result of your control solution test to the range
printed on your test strip vial label for the control solution
level that you used. Your control solution result should fall
within this range.
Example Result:
If Your Control Solution Test Results Are Out of the Range:
• Repeat the test and refer to Chapter 13 of this guide to see
potential causes of error.
• If the results from the control solution tests continue to
fall outside that range, do not use the meter to test your
blood glucose.
• Call Customer Service. The phone number is printed at
the start of this guide and also printed on the back of your
meter.
Level 2: 80-125 mg/dL
Level 4: 220-320 mg/dL
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.