[3] Applying Control Solution:
1. Shake the control solution bottle.
2. Discard the first drop of control solution.
3. Wipe the bottle tip clean using a clean tissue or paper
towel.
4. Dispense a second drop onto a clean surface, such as an
unused resealable plastic bag.
5. Avoid handling the test strip's sample area (where con-
trol solution should be applied).
6. Bring the tip of the test strip to the control solution
sample at a 90 degree angle immediately so that the
control solution does not evaporate.
The meter will automatically detect if a test is performed with
control solution.
Apply Sample to Test Strip Animation:
This is the visual
fi ll window. Do
not put Control
Solution here.
Apply control
solution here.
27