Lsm 880 With Airyscan On Large System Table - Zeiss LSM 880 Notes On Device Safety And Installation Requirements

Hide thumbs Also See for LSM 880:
Table of Contents

Advertisement

LSM 880
1.2

LSM 880 with Airyscan on Large System Table

Fig. 2
Space requirements for LSM with Airyscan on large system table
(measurements in mm)
The realtime PC and the liquid cooling system are located beneath the system table. The Argon
laser power supply is placed on top of the laser rack. The user PC is located under the computer
table. If present, the laser module UV is located on top of the Laser Rack of the LSM 880. The total
height of these racks is then 1 m. For systems with Axio Observer rear port the size of the system
table is 900 mm wide and 1200 mm deep. For systems with Axio Imager the size of the system
table is 900 mm wide and 750 mm deep.
11/2014 V_05
CHAPTER 2 - SETUP REQUIREMENTS
Space Requirements
000000-2071-466
ZEISS
39