Snap-On ControlTech Micro Calibration Procedure page 6

Electronic torque-angle wrench
Hide thumbs Also See for ControlTech Micro:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 18
2. Scroll to the CONFIGURE menu selection by pressing the Up key two times:
CONFIGURE
EXIT
SET HEAD LENGTH
3. Press the Enter key to display the Configure menu:
EXIT
MODE SETUP
PRESET LOCK
Note: The Wrench real-time-clock Date is stored in non-volatile memory during
torque calibration. The stored date is used to calculate the next calibration
interval. If the clock has already been set, skip to step 12.
4. Scroll to the SET DATE/TIME menu selection using the Up/Down keys:
JOB MODE
CALIBRATION
SET DATE/TIME
5. Press the Enter key to display the Set Date screen:
SET DATE
2017/01/01
YR MON DAY
6. Set the year by incrementing or decrementing the value using the Up/Down keys
then press the Enter key to accept the year and highlight the month value.
7. Set the month by incrementing or decrementing the value using the Up/Down keys
then press the Enter key to accept the month and highlight the day value.
8. Set the day by incrementing or decrementing the value using the Up/Down keys
then press the Enter key to accept the day and display the Set Time screen:
SET TIME
12
: 30 : 00
HR MIN SEC
9. Set the hour by incrementing or decrementing the value using the Up/Down keys
then press the Enter key to accept the hour and highlight the minutes value.
10. Set the minutes by incrementing or decrementing the value using the Up/Down
keys then press the Enter key to accept the minutes and highlight the seconds
value.
11. Set the seconds by incrementing or decrementing the value using the Up/Down
keys then press the Enter key to accept the seconds and display the Configure
screen.
12. Scroll to the CALIBRATION menu selection:
JOB MODE
CALIBRATION
SET DATE/TIME
13. Press the Enter key to display the Password Entry screen:
ENTER PASSWD
9 1 7 4
14. Enter the password: 91748 by pressing the Up or Down keys to
increment/decrement the highlighted digit, followed by the Enter key after each digit
is set.
15. Following the entry of the last password digit, the Calibration menu will be
displayed:
EXIT
CAL TORQUE
CAL ANGLE
8
ControlTech
4
TM
Micro Wrench