Flight Medical Innovations 60 Series Service Manual page 98

Table of Contents

Advertisement

Troubleshooting Guide
Error Messages
Displayed error
HIGH MV INS
HIGH MV EXH
HIGH / LOW P BASE
PCV NOT REACHED
LOW MV EXH
MOTOR FAULT
PRESS SENSOR
POWER FAULT
MEMORY FAULT
MAIN BAT CHARGER
MAIN BAT V HIGH / LOW
9-4 | FLIGHT 60 Service Manual
Potential Cause
Air leakage
Calibration problem
Hardware problem
Solenoid malfunction
Calibration problem
Hardware problem
Air leak
Hardware problem
Motor control
Motor
Patient pressure sensor higher
than outlet
Check log file to identify failed
power:
3.3V POWER FAULT
MAIN BOARD 10V POWER FAULT
POWER BOARD 10V POWER
FAULT
5V POWER FAULT
Parameters mismatch
Charger malfunction
Out of range
Suggested Action
Check internal hoses connections
Check internal hoses connections
Perform Volume factor calibration
Replace Main Board
Replace manifold
Check solenoid
Perform Circuit Test
Replace solenoid
Perform Volume Factor calibration
Replace manifold
Check internal hoses connections
Replace Main Board
Replace Power Board
Replace manifold
Perform Pressure sensors calibration
Replace Main Board
Replace Main Board
Check cables
Replace Power Board
Replace Main Board
Check cables
Replace Power Board
Replace Main Board
Perform OVP
Replace Main Board
Replace battery
Replace Power Board
Replace battery
Replace Power Board