Removing and Reinstalling Modules
Replacing the Main Board
Figure 36 – Disconnecting LS2 connector
5. Disconnect the P13 flat cable by pressing hard on the locker and then pulling it
out.
Figure 37 – Disconnecting the P13 Flat Cable
6. Remove the six Philips screws securing the Main Board to the front panel.
Flight 60 Service Manual | 4-19