Flight Medical Innovations 60 Series Service Manual page 50

Table of Contents

Advertisement

Removing and Reinstalling Modules
Replacing the Main Board
Figure 38 – Detaching the Main Board from the Front Panel
7. Remove the grounding screw at the top left of the board.
8. Lift the Main Board. Carefully detach the flat cables from the connectors on the
main board.
9. Remove the J3 flat cable by clicking the locker on the main board and then pulling
out the cable.
4-20 | FLIGHT 60 Service Manual
Figure 39 – Detaching the Grounding Screw