Calibration and Verification tests
Calibration
8. Press Calibrate Pressure and then Enter /
pressure.
9. The following screen opens:
10. Check the Patient Pressure and Outlet Pressure values, verify the following:
Values match the pressure measurement of the external pressure meter.
Values are between 50 and 75 cmH
The difference between the Patient Pressure and Outlet Pressure values is no
more than 0.1.
If one of the conditions above is not met, calibrate the pressure sensors:
1. If the pressure on the external pressure meter is in the 50-75 cmH
range, continue to step 2.
If the pressure is less than 50 cmH
Press Cancel /
Press Calibrate Pressure again. The calibration screen will come up with the
following message " You can set the solenoid and motor commands for
pressure sensor" and displays default values of 65 200.
Use the UP /
initial value of 65) or Manual Breath /
motor command (initial value of 200) and press Enter/
Verify that the pressure on the external pressure meter is in the 50-75 cmH
range. If not then repeat the above steps until the pressure is the correct
range.
Continue to step 2.
7-8 | FLIGHT 60 Service Manual