Calibration and Verification tests
Calibration
8. Press Calibrate Pressure and then Enter /
pressure.
9. The following screen opens:
10. Check the Patient Pressure and Outlet Pressure values, verify the following:
 Values match the pressure measurement of the external pressure meter.
 Values are between 50 and 75 cmH
 The difference between the Patient Pressure and Outlet Pressure values is no
more than 0.1.
If one of the conditions above is not met, calibrate the pressure sensors:
1. If the pressure on the external pressure meter is in the 50-75 cmH
range, continue to step 2.
If the pressure is less than 50 cmH
Press Cancel /
Press Calibrate Pressure again. The calibration screen will come up with the
following message " You can set the solenoid and motor commands for
pressure sensor" and displays default values of 65 200.
Use the UP /
initial value of 65) or Manual Breath /
motor command (initial value of 200) and press Enter/
Verify that the pressure on the external pressure meter is in the 50-75 cmH
range. If not then repeat the above steps until the pressure is the correct
range.
Continue to step 2.
7-8 | FLIGHT 60 Service Manual
2
O:
2
.
and DOWN /
buttons to change the solenoid command (
. The ventilator starts to build
O.
and Panel lock /
O
2
buttons to change
.
O
2