13. Disconnect the silicon tubes from the solenoid.
14. Release the Manifold power cord from the wire saddles.
15. Remove the six Torx screws from the manifold base (three on each side).
16. Carefully lift out the manifold assembly.
Don't apply any force on the piston drive mechanism. Don't hold the manifold
by the bearings or by the metal rods.
Figure 48 – Disconnecting the Flex Tubes
Figure 49 – Detaching the Manifold Base
Removing and Reinstalling Modules
Replacing the Manifold Assembly
Flight 60 Service Manual | 4-25