Removing and Reinstalling Modules
Replacing the Front Panel Assembly
For iO
Internal Mixer model
2
6. Disconnect the 2 silicon tubes from the O
7. Disconnect the nebulizer cable from the O
8. Disconnect the mixer cable from the O
9. Disconnect the O
10. Disconnect the 3 silicon tubes coming from the main board. Note the color coding.
Disconnect them from the Y connectors and not from the board itself.
11. Disconnect the solenoid board flat cable.
4-16 | FLIGHT 60 Service Manual
Figure 32 – Torx Screws on Front Panel Bracket
leak sensor cable from the O
2
Figure 33 – Disconnecting from O
mixer neb board.
2
mixer neb board.
2
mixer neb board.
2
mixer neb board.
2
mixer neb board
2