Fluorescein; Rhodamine - Valeport Hyperion Operating Manual

Fluorometer
Table of Contents

Advertisement

3.1.3. Fluorescein

Performance
Excitation:
470 nm
Detection:
545 nm
Dynamic Range: 0-500 ppb
2 gain settings: 0-25 and 0-500 (software controlled)
Instrument
<0.01 ppb*
Detection limit:
Actual
0.03 ppb**
Detection limit:
Linearity:
0.99 R
Response Time: 0.03 to 2 sec
* 3x SD in RO water
** Calibrated against Fluorescein solution

3.1.4. Rhodamine

Performance
Excitation:
520 nm
Detection:
650 nm
Dynamic Range: 0-1000 ppb
2 gain settings, 0-50, 0-1000 (software controlled)
Instrument
<0.01 ppb*
Detection limit:
Actual
0.06 ppb**
Detection limit:
Linearity:
0.99 R
Response Time: 0.03 to 2 s
* 3x SD in RO water
** Calibrated against Rhodamine solution