Electrical; Connector Pin-Out - Valeport Hyperion Operating Manual

Fluorometer
Table of Contents

Advertisement

0901814c - Hyperion Fluorometer

6. Electrical

Voltage:
9 - 30V DC isolated
Power:
40mA @ 12V DC

6.1. Connector Pin-Out

Pin
Function
1
RS232 GND
2
Tx 232 out from SVS (485A)
3
Rx 232 in to IPS (485B)
4
+V(8-20V)
5
/Enable 485
6
-V
Note: to enable RS485 comms, link pins 1 and 5.