10.2
X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the VEX-P300.
The X-ray doses of the VEX-P300 in the test report were measured in accordance with
the IEC collateral standards. The VEX-P300 was designed in accordance with Part 1.
General Requirements for Safety, IEC 60601-1-3.
Tube Model Name
Generator Model Name
X-ray Dose Table
10.2.1
Instrument
Dose Meter
Dose Table (65 kVp, 2.5 mA, FOV: Ø 6 cm, SSD 200 mm, at Al 6 mm)
EzRay Air (Model: VEX-P300) User Manual
Test Condition
Model Name
Loading Factor
Test Equipment
Manufacturer
Piranha
t (s)
0.05
0.08
0.1
0.15
0.2
0.25
0.3
0.45
0.5
10. Appendix
VEX-P300
V1-650304
DG-S0101V1 (Inverter type)
65 kV, 2.5 mA
Model
S/N
255
CB2-0750141
Dose (μGy)
50.9
91.2
117.9
185.1
252.1
319.2
386.2
587.4
654.4
77