6. Troubleshooting

Troubleshooting
6.
In instances of abnormal operation, error messages will be displayed on the Control
Panel. If a problem persists, please request assistance from the customer support
information services.

Alarm/Error Messages

Error Code
E.02
E.03
E.04
E.05
A.06
A.07
A.08
A.09
A.10
A.11
E.12
60
A.0X: A problem occurred, and the system performs the correction
automatically. This alarm clears after the correction is completed.
E.0X: An error occurred. Turn the power off, and then turn it back on. If
the error persists, contact your Service Representative.
Check Parameter
X-ray Generator
X-ray Generator
System
User
Battery
Description
Error related to X-ray exposure is not possible to
exposure X-ray in the state while "E.02", "E.03",
"E.04", "E.05" where power is maintained.
After X-ray exposure related error occurs, when
the equipment is turned off and turned on, X-ray
exposure is normally performed.
Appears when the system needs cooling time
due to continuous operation. This alarm clears
when the system temperature goes down to
normal.
Appears when the Remote Exposure Switch has
been held down. Release the switch.
Appears when the exposure button has been
pressed and released before the X-ray exposure
finished with the default exposure time. Press
and hold the exposure button for the duration of
the exposure time.
Appears when the battery voltage is higher than
the reference value. Check the battery.
Appears when the battery voltage is lower than
the reference value during X-ray exposure.
Charge the battery.
Appears when the battery is being charged.
Operate the system after disconnecting the
battery charger from the system.
Appears when battery level is below the
reference value while X-ray exposure.