12. X-Ray Dose Data; X-Ray Dose Table - Vatech EzRay Premium VEX-S100W User Manual

Table of Contents

Advertisement

12. X-Ray Dose Data

The X-ray dose data is extracted from the X-Ray Dose Test Report for the EzRay
Premium. The X-ray doses of EzRay Premium in the test report were measured in
accordance with the IEC collateral standards. EzRay Premium was designed in
accordance with Part 1. General Requirements for Safety, IEC 60601-1-3.
Tube type / Inverter type
12.1

X-Ray Dose Table

Instrument
Dose Meter
t (s)
0.04
0.05
0.08
0.10
0.15
0.20
0.25
0.30
0.45
EzRay Premium User Manual
Test Condition
Brand Name (Model)
Tube type
Loading Factor
Test Equipment
Manufacturer
Piranha
Dose Table
Dose (mGy)
0.05
0.07
0.10
0.13
0.19
0.25
0.31
0.38
0.57
12. X-Ray Dose Data
EzRay Premium (VEX-S100W)
Toshiba D-041SB
60 kV, 6.0 mA
DG-10A05T3 / INV-05
Model
S/N
255
CB2-08050121
t (s)
Dose (mGy)
0.50
0.59
0.55
0.70
0.60
0.76
0.70
0.89
0.80
1.01
0.90
1.15
1.00
1.29
1.50
1.93
2.00
2.57
47

Hide quick links: