8. Control Panel Operation Guide

The Control Panel of the power unit is used to make adjustments to the mattress and also indicates
fault conditions/service requirements. These are either visual (indicator lights) or audible.
1
1
Comfort Control Dial
2
Control Panel Lockout Button
3
Control Panel Lockout Indicator
4
Service Indicator
5
Power Failure Alarm Indicator
6
Low Pressure Alarm Indicator
8.1

Comfort level

The comfort level adjustment dial allows carers to adjust pressures within a safe pre-set range to
provide patients with enhanced comfort or support whilst maintaining a very good level of protection
and therapy. Qualified clinical advice must always be taken before adjusting mattress overlay
pressures.
When the dial is turned towards Firm, the output pressure will increase to provide a higher pressure
output and thus increased support.
When the dial is turned towards Soft, the output pressure will be decreased to provide a lower
pressure output and thus increased comfort.
To check if the pressure is adequately supporting the patient, slide one hand between the mattress
overlay and base mattress to feel under the patient's bottom. You should be able to slide the hand in-
between and an acceptable range is approximately 25 to 40 mm (1" to 1-1/2") to ensure the patient Is
not bottoming out.
Issue: 5 - 13/02/2019
12
11
10
9
8
2
3
4
5
7
Alarm Mute Button
8
Power Indicator
9
Normal Pressure Indicator
10
Static Mode Indicator
11
Mode Selection Button
12
Cycle Time Indicators
8
8
6
7
Page 11 of 24