8.2
Control panel lockout
If the control panel is not used for a period of 30 seconds it will lock out and inhibit the
use of the functions. Additionally, if you wish to lock out the control panel press the Lock
Out button and all functions will be locked. This is to prevent the system being altered
accidently.
In order to unlock the control panel, simply press and hold the Lock Out button for 3
seconds, the control panel will now be active for use.
8.3
Function mode switch
8.3.1 Alternate
Alternating mode is the default mode for the system. Within this mode the mattress will
operate in an alternating 1-in-2 cell cycle. The alternating cycle will continue at the
selected cycle time until another mode is selected.
8.3.2 Static
Pressing the Mode button until the Static indicator illuminates puts the system into
Static mode.
Within this mode the mattress will maintain a selected constant pressure. After 30
minutes the system will automatically revert back to Alternating mode.
8.4
Alternate cycle time selection
The alternating cycle times can be selected to provide an individualised care program for
each patient.
Within Alternating mode the cycle time can be selected by pressing the Mode button.
Selections can be made from 10-20 minutes at 5 minute intervals. The cycle time
selection will be indicated by the relevant indicator light being illuminated.
8.5
Alarm mute
The Alarm Mute button temporarily resets the audible Low Pressure/Power
Failure/Service alarms. Should the situation not be resolved and the fault condition
continues the alarm will resume notifying carer.