Ultimate Healthcare Heritage II Digital Turn Quick Reference Manual

Ultimate Healthcare Heritage II Digital Turn Quick Reference Manual

Alternating mattress replacement system with a cell on cell construction
Hide thumbs Also See for Heritage II Digital Turn:

Advertisement

Quick Links

Heritage II Digital Turn
Quick Reference Guide
Please refer to the product User Manual for full instructions for use.
30
Control Panel Lock Out
To unlock panel:
Press & hold for 3 seconds to unlock control panel.
To lock panel:
To lock out the control panel press the Lock Out
button and all functions will be locked.
If the control panel is not used for a period of 30
seconds it will automatically lock out and inhibit
the use of the functions.
Comfort Control
The Soft and Firm buttons allow adjustment of
When pressing the Firm button, the output pressure
will increase to provide a higher pressure output
and thus increased support.
When pressing the Soft button, the output
pressure will be decreased to provide a lower
pressure output and thus increased comfort.
Auto-Detection -
detecting patient weight (distribution)
Simultaneously pressing the Soft and Firm buttons on the
control panel will manually activate Auto-Detection.
will start detecting the patients weight
internal pressures appropriately. The system will
continuously check this every 2 hours.
Auto Firm Mode
Pressing the Auto Firm
cells to a single pressure setting.
The system will automatically return to
Last Mode Of Operation after 20 minutes.
Function Mode button
Pressing the Mode button allows you to cycle
between the different modes available such as
Alternating Static
A Full Risk Assessment should be completed before using this System
255
kg
165 205
130
100
75
50
to set
Constant Low Pressure
-Please refer to User Manual for guidance
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
CLP
Mode
Alternating Mode
Alternating mode is the default mode for the
The alternating cycle will continue at the selected
cycle time until another mode is selected.
Static Mode
Pressing the Mode button until the Static indicator
illuminates puts the system into Static mode.
The system will automatically return to
Last Mode Of Operation after 20 minutes.
Pressing the Mode button until the CLP indicator
illuminates puts the system into Constant Low
Pressure mode.
Alternating Cycle Time Selection
To select press Cycle button until required cycle
time is displayed in LCD screen.
Alarms
Alarm Mute
The Alarm Mute button temporarily resets the
audible Low Pressure/Power Failure/Service
alarms. Should the situation not be resolved and
the fault condition continues the alarm will resume.
Low Pressure Alarm
Upon detection of low pressure, an audible alarm will be
heard and the Low Pressure indicator will illuminate.
Power Failure Alarm
If mains power is removed from power unit or the power cable
is unplugged without turning the power unit OFF, an audible
alarm will sound and the Power Failure indicator will illuminate.
Service Alarm
Should your system develop a fault condition whilst in use,
an audible alarm will be heard and the Service indicator will
illuminate.
Mode

Advertisement

loading

Summary of Contents for Ultimate Healthcare Heritage II Digital Turn

  • Page 1 Heritage II Digital Turn Quick Reference Guide Please refer to the product User Manual for full instructions for use. 165 205 Mode Alternating Mode Control Panel Lock Out • Alternating mode is the default mode for the To unlock panel: •...
  • Page 2 Mattress Setup Secure the mattress to the moving parts of the bedframe using Cable Management System Cable management fixings are the securing straps, and when the system is fully inflated it is located on each side of the mattress underneath the flap of the ready for the patient to be positioned onto the mattress.