13. Cleaning and Decontamination

The following processes are recommended, but should be adapted to comply with the local or
national guidelines (Decontamination of Medical Devices) which may apply within the Healthcare
Facility of use.
The Tamora Plus II system should be routinely decontaminated between patients and at regular
intervals while in use; as is good practice for all reusable medical devices.
Warning: Disconnect the power unit from the electricity supply before carrying out
cleaning/decontamination procedures.
Do not immerse or soak power unit.

13.1 Basic cleaning information

Caution: Only use disinfectants designed for cleaning healthcare equipment i.e.
Sodium Hypochlorite or similar (up to 10,000 ppm available chlorine).
Caution: Do not use abrasives (scouring powder), scourers or other
materials/agents which could damage the mattress system.
Caution: Do not use sodium carbonate or phenol based solutions.
Caution: Do not use fabric softener or non-biological washing detergent.
Caution: Do not immerse the power unit in water.
Caution: When cleaning/disinfecting, ensure that only a damp cloth is used.
Caution: After cleaning, dry the mattress out of direct exposure to sunlight.
Caution: Using inappropriate detergents or disinfectants and not observing the
manufacturer's guidelines may result in damage to the mattress which Ultimate
Healthcare cannot be held liable for.
Caution: The appropriate qualified staff must be consulted when specifying a
suitable cleaning fluid. Ultimate Healthcare shall not be liable for any damages
caused by the use of inappropriate detergents or disinfectants.

13.2 Mattress and cover disinfection

The Tamora Plus II mattress, mattress cover and air pipe cover can be cleaned using the following
simple procedures in accordance with your Local Infection Control Policy:
Liberally swabbing with a damp cloth pre-soaked with hot water at 84°C containing detergent,
and then drying.
Swabbing with a solution of sodium hypochlorite (up to 10,000 parts per million available
chlorine) and then drying.
Caution: Frequent or prolonged exposure to higher concentration disinfectant
solutions may prematurely age the fabric cover of mattresses. Surfaces must be
protected during use and rinsed and thoroughly dried after application of a
disinfectant.