Carbon Dioxide Detector; Unpacking The Model 9847 - Nonin 9847 Operator's Manual

Handheld pulse oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

General Description

Carbon Dioxide Detector

9847 determines approximate CO
changes in the airway of intubated patients by measuring the
2
absorption of mid-infrared light passed through the airway adapter tube. The approximate CO
2
concentration change is indicated by an 8-segment LED bar graph display. The CO
detector relies on
2
the assumption that the inhaled air contains minimal amounts of CO
.
2
Breaths are indicated when the CO
level increases by approximately 5 mmHg during exhalation. A
2
detected breath is indicated on the CO
bar graph and by an audible breath beep.
2
When no breath is detected:
• a high priority patient audible alarm sounds,
• the no breath indicator blinks,
• the lowest CO
bar graph segment will be illuminated, and CO
readings will be displayed with the
2
2
next breath detected.
Medium priority (equipment) alarms occur when:
• the CO
sensor is unplugged,
2
• the airway adapter tube is removed from the CO
sensor,
2
• the light path is blocked, or
• a CO
sensor failure occurs.
2
A high priority (patient) alarm overrides a medium priority alarm. If the audible alarms are disabled, the
third and sixth bars will be solidly illuminated.

Unpacking the Model 9847

Contact the carrier immediately if the shipping carton is damaged. Carefully unpack the device and its
accessories. Nonin's standard package configuration consists of the following items:
• 1 Model 9847 Pulse Oximeter and CO
Indicator
2
• 1 Operator's Manual (on CD)
• 1 Model 8000AA-1 Adult Articulated Finger Clip Sensor
• 1 Model 9840SA Carbon Dioxide Sensor
• 3 Model 9840AAT Airway Adapter Tubes
• 6 AA-Size Alkaline Batteries
If any item on this list is missing or damaged, contact your distributor.