Pulse Oximeter Startup - Nonin 9847 Operator's Manual

Handheld pulse oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

Using the 9847 Pulse Oximeter
• The %SpO
and pulse rate
2
1. "888 888".
2. The current time (if set) or "00 00" if the time is not set.
3. Display 3 software revision numbers; oximetry/display software rev #, then CO
memory software rev #, then sound module software rev. # approx 1 second each.
4. Any sensor connections issues:
if a sensor is not connected the display will revert to blank, or
if a sensor is connected but is not detecting an adequate signal, a single dash (-) will appear in
the middle position of both displays, or
if entering setup mode, the device will display "ALr" and "dFt".
The CO
bar graph sequence is:
2
• each segment illuminates once,
• remains blank (until CO
• only the bottom segment illuminates when ready for use if CO
• The yellow audible alarm disabled indicator is lit until initialization is complete
and a CO
or pulse oximeter sensor is connected.
2
NOTE: The audible alarm disabled indication cannot be turned off until a sensor
is plugged in.
NOTE: When 9847 is turned on and the alarms are disabled, the yellow audible alarm disabled
indicator will remain illuminated.

Pulse Oximeter Startup

Apply the pulse oximeter sensor to the patient as directed in the sensor Instructions for Use. Verify
operation by:
• the pulse quality indicator
• the pulse rate
and SpO
• the pulse quality indicator blinking is correlated to the pulse rate for at least 10 seconds.
If the pulse quality indicator LED is blinking red or yellow or is blinking inconsistently, reposition the
sensor or replace the sensor.
displays will sequence as follows.
sensor is connected to the device), or
2
is blinking green,
(%SpO
) displays are showing values, and
2
2
2
sensor is connected to the device.
2
Audible
Alarm
Disabled
Indicator
(Yellow)
%SpO 2
8 8 8
8 8 8
15