Section 5 – Troubleshooting
Local Panel and Remote Keypad Message Types
The local panel displays two types of messages:
Status Messages – Provide information and require no operator action.
Error Messages – Indicate an error condition inside the imager. Some
messages tell you what action to take to fix the error, but in some cases a
service call is needed to deal with a problem. If the error message does not clear
after you perform the specified action, call for service.
Table 5-1 lists all messages that can appear on the two lines of the local panel
display:
The Information Line (line 1) of each display contains the error title and (in most
cases) a two-digit error code.
The Action Line (line 2) contains additional information/procedures about the
error condition.
Note
The keypad indicates that an error has occurred by flashing the Check Imager
indicator, which refers the operator to the imager's local panel for more detailed
error information.
Local Panel Messages
If the Operator Action specified in Table 5-1 does not correct the machine problem,
call for service (see page 5-20).
Table 5-1. Operator Related Local Panel Messages
Message
ASIC LINE BUFFER FAIL (48)
CALL SERVICE – PRESS ENTER TO
CONTINUE
ASIC SELF TEST FAIL (60)
CALL SERVICE – PRESS ENTER TO
CONTINUE
2000 March Rev. B
8186546
Section 5 – Troubleshooting
Operator Action
Power the imager off, wait 10-20
seconds, then power back on. If
message reappears, call for service.
Power the imager off, wait 10-20
seconds, then power back on. If
message reappears, call for service.
5-1