Agilent Technologies 54753A Programmer's Reference Manual page 34

Tdr plug-in modules
Table of Contents

Advertisement

Command
<source>
<number>
Command
<source>
<number>
Command
<source>
<number>
Measure Commands
:MEASure:VAMPlitude[?] [<source>]
Calculates the difference between the top and base voltage of the specified
source.
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
:MEASure:VAVerage[?] {CYCLe | DISPlay},[<source>]
Calculates the average voltage over the displayed waveform.
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
:MEASure:VBASe[?] [<source>]
Measures the statistical base of the waveform.
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
32

Advertisement

Table of Contents
loading

This manual is also suitable for:

54754a

Table of Contents