Agilent Technologies 54753A Programmer's Reference Manual page 28

Tdr plug-in modules
Table of Contents

Advertisement

Command
<source>
<number>
Command
<source>
<number>
Measure Commands
:MEASure:FALLtime[?] [<source>]
Measures the time at the upper threshold of the falling edge, measures the
time at the lower threshold of the falling edge, then calculates the fall time.
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
:MEASure:FREQuency[?] [<source>]
Measures the frequency of the first complete cycle on the screen using the
mid-threshold levels of the waveform (50% levels if standard measurements
are selected).
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.