Acquire Commands

Command
:ACQuire:BEST[?] {THRuput | FLATness}
When averaging is enabled with ACQuire:AVERage, the FLATness option
improves the step flatness by using a signal processing algorithm within the
instrument. You should use this option when performing TDR measurements
or when step flatness is important. The THRuput option improves the
instrument's throughput and should be used whenever best flatness is not
required.