Reference Manual
00809-0100-2460, Rev DE
LED board
6. Unplug the Main Board connector.
7. Remove the Power Supply Unit and replace it with a new one.
8. Tighten the screws, attach the Main Board connector, and plug in the power connection.
9. If removed, replace the lid on the Rosemount 2460 housing and ensure that the lid is secured with the
ring.
10.Ensure that the lid is properly closed and sealed.
Service and Troubleshooting
Service and Troubleshooting
Main Board connector
October 2018
113