8.2.5 Remove the two white nylon elbows from each of the orange pump tubes.
8.2.6 Remove the orange pump tube from the pump roller and bracket.
8.2.7 Install the new orange pump tube into the bracket and onto the roller.
8.2.8 Insert the two nylon elbows into the new pump tube.
8.2.9 Turn the unit ON and replace the front panel.

8.3 Register New Pump Replacement Tube Kit

8.3.1 From the Main Menu, Press [ 2 ] on the key pad to select SET-UP.
8.3.2 Press [ 2 ] on the key pad to select Consumables
8.3.3 Press [ 2 ] on the key pad to select Review Tubing
8.3.4 Press [ 1 ] on the key pad to select Change Kit
8.3.5 Enter the Kit Serial Number and press [ # ] on the key pad to Accept
8.3.6 Enter the Kit Batch Number and press [ # ] on the key pad to Accept
8.3.7 Press [ * ] on the key pad once
8.3.8 Press [ 1 ] on the key pad to Update Tubing Log
8.3.9 Press [ * ] on the key pad to return to the Main Menu
8.3.10 Press [ 1 ] on the key pad to select Run
8.3.11 Press [ 3 ] on the key pad to select Prime
The unit is now ready to run samples.