Figure 4-4: Camera Angled Position for Stabilized Binocular
and Monocular Recording

4.1.1 Cabling for the Standard Camera

After the camera is mounted onto the Desktop Mount, connect the two illuminator cables that
come out of the Desktop mount to the left side of the EyeLink 1000 high-speed camera.
Connect the EyeLink 1000 power supply that was provided with your system to the power
connector on the left side of the camera (see Figure 4-5).
Figure 4-5: Camera and Illuminator Cables from Desktop
Mount

4.1.2 Cabling for the Fiber Optic Camera

After the Camera Adapter with the Camera Head and lens are mounted onto the Desktop
Mount, connect the two illuminator plugs emerging from the mount into the 3.5 mm (1/8")
mini-plug sockets of the Integrated Extension Cable.
Desktop Mount / EyeLink Remote Installation