Data Input Format

You can connect a computer to your balance/scale to
send commands via the balance/scale interface port
to control balance/scale functions and applications.
A control command can have up to 4 characters. Each
character must be transmitted according to the settings
configured in the Setup menu for data transmission.
Format for Control Commands
Format:
Esc
!
CR
Esc:
Escape
!:
Command char.
CR:
Carriage return (optional)
LF:
Line feed (optional)
Command character ! Meaning
K
L
M
N
O
P
R
S
T
U
V
W
LF
Weighing mode 1
Weighing mode 2
Weighing mode 3
Weighing mode 4
Block keys
Print
Release keys
Restart
Tare and zero (combined)
Tare ("tare only")
Zero
External calibration/adjustment
(depends on menu setting)