Additional Characteristics - Keithley System SourceMeter 2601B Instrument Specifications

Hide thumbs Also See for System SourceMeter 2601B:
Table of Contents

Advertisement

Models 2601B, 2602B and 2604B System SourceMeter
ADDITIONAL MEASUREMENT CHARACTERISTICS
Contact check
13,14
specifications
Current measure settling
15
time
Input impedance

ADDITIONAL CHARACTERISTICS

Maximum load impedance
Common mode voltage
Common mode isolation
Sense high input impedance
Maximum sense lead
resistance
Overrange
13
Includes measurement of SENSE HI to HI and SENSE LO to LO contact resistances.
14
Contact check is not available with the Model 2604B.
15
Compliance equal to 100 mA
Specifications are subject to change without notice
6 of 13
®
Instrument Specifications
Maximum measurement time
Speed
to memory for 60 Hz (50 Hz)
Fast
1.1 ms (1.2 ms)
Medium
4.1 ms (5 ms)
Slow
36 ms (42 ms)
Time required to reach within 0.1 % of final value after source level command is
processed on a fixed range
Values below for Vout = 1 V
Current range
1 mA
> 10 GΩ
Normal mode
10 nF
250 V DC
> 1 GΩ
< 4500 pF
> 10 GΩ
1 kΩ for rated accuracy
101 % of source range
102 % of measure range
Accuracy (1 year)
23° C ± 5° C
± ( % reading + ohms)
5 % + 10 Ω
5 % + 1 Ω
5 % + 0.3 Ω
Settling time
< 100 µs
High-capacitance mode
50 µF
SPEC-2602B Rev. B / February 2014

Advertisement

Table of Contents
loading

This manual is also suitable for:

System sourcemeter 2602bSystem sourcemeter 2604b

Table of Contents