The analyzer stores a maximum of 18,000 data points. After 18,000 data points are stored, the oldest
data points are overwritten with new data.
1. Select VIEW DATA.
2. Select an option.
Option
ANALYZER DATA
MEASUREMENT DATA Shows measurement information (refer to
LOG DATA
Element
CELL TEMP
REAGENT TEMP
AMBIENT TEMP
SAMPLE TEMP
AIR PRESS
LED DUTY CYCLE
HEATER DUTY CYCLE Percentage of time the colorimeter heater is on to maintain a constant 50 °C (122 °F)
SAMPLE FLOW
SAMPLE PRESS 1
SAMPLE PRESS 2
REAGENT 1
REAGENT 2
REAGENT 3
STD SOLUTION
FAN SPEED
LEAK COUNTS
Element
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME
Description
Shows analyzer status information (refer to
Selects the data log and/or the event log. DATA LOG—shows the measurement
values. Select start time, number of hours and/or number of readings. EVENT LOG
—shows all analyzer information (e.g. alarms, warnings, configuration changes,
and so on.). Select start time, number of hours and/or number of readings.
Table 2 Analyzer data
Definition
Temperature of the colorimeter cell block heater (ideally 49.8 °C to 50.2 °C (121.64 °F
to 122.36 °F))
Reagent temperature before it enters the colorimeter
Air temperature inside the electronics area
Sample pre-heater block temperature (typically 45 °C to 55 °C (113 °F to 131 °F) but
could reach 58 °C (136.4 °F))
Reagent air pressure in the reagent bottles (ideally 3.95 to 4.10 psi)
Depends on condition of colorimeter cell and age of analyzer (typically 7,200 to
40,000 counts)
Approximate sample flow into the colorimeter, measured during the flush cycle
Sample pressure before the pre-heater block (ideally 2 to 4.5 psi depending on
incoming sample pressure)
Sample pressure after the sample pre-heater, which is used to calculate the sample
flow. The sample pressure is almost zero when the flush is set to off and is
approximately 0.2 psi when the flush is on (depends on inlet pressure and flow).
Level of reagent remaining
Level of reagent remaining
Level of reagent remaining
Level of standard solution remaining
The speed of the ventilation fan.
Indication of possible fluid leaks (range 0 to 1023). A count of over 511 indicates a fluid
leak
Table 3 Measurement data
Definition
Time of last measurement.
Table
2).
Table
3).
English 15