Recommended Initial Testing Procedure

SpHb Spot Check Measurement
Note: Potentially inaccurate results can occur if the user does not properly
follow the testing procedure. As with any diagnostic test, clinicians should
always evaluate hemoglobin measurements in the perspective of the clinical
context of the patient.
Sensor Site Selection
1.
Assess the patient's arm of the selected site for any restrictions from
shoulder to fingertip.
2.
Restrictive garments or accessories should be removed as it can
impede blood flow to the sensor site. (Example: purse, backpacks,
watches, jewelry and blood pressure cuff).
Do not use sites with any of the following conditions:
An anatomically abnormal finger (e.g. damaged, clubbed,
deviated, etc.)
A finger or arm that has experienced previous surgical
procedures
A finger or arm that is currently receiving an IV infusion
3.
Site should be cleaned of debris and dry prior to testing.
4.
Select the patient's testing finger in the following priority:
Non-dominant ring or middle finger
Dominant ring or middle finger
www.masimo.com
Chapter 4: Operation