Modes
Pronto-7 is equipped with five modes; three sensitivity modes and two test
modes. The sensitivity modes include normal, max and multi. The test modes
include SpO
only and Measure through Motion, identified on the Pronto-7
2
screen as measure under Low SIQ.
If any of the three sensitivity modes are toggled to On, the Pronto-7 will remain
in the test mode until the user manually configures the mode to Off, or if SpO
only test mode is turned On, as this test mode overrides all sensitivity modes.
Measure through Motion test mode can be turned on and used simultaneously
with any of the four other modes. If either of the two test modes is toggled to
On, the Pronto-7 will remain in this test mode until the user manually
configures the mode to off.
All modes can be accessed from the Main Menu icon. Some modes can be
accessed in additional ways.
Sensitivity Modes Details
Normal is the default sensitivity mode for the Pronto-7.
Max sensitivity mode should provide fewer "Test Incomplete"
messages compared to normal sensitivity mode.
Multi test mode provides an averaged result based upon three tests.
Pronto-7 will perform multiple individual tests until three quality
samples are obtained which will be averaged for the single test
result.
Test Modes Details
SpO
only test mode displays SpO
2
does not consume a spot check test. If SpHb is needed after the
SpO
only mode test is complete the user can access SpHb, which
2
will deduct a spot check test. See Accessing SpHb in SpO2 Only
Test Mode on page
If SpO
only test mode is on, the Pronto-7 will remain in this test
2
mode until the user manually configures the mode to off. See
Accessing Sensitivity and Test Modes on page
Measure through Motion mode allows for fewer incomplete tests and
can be enabled from the Test Mode icon.
If Measure through Motion test mode is on, the Pronto-7 will remain
in this test mode until the user manually configures the mode to off.
See Accessing Sensitivity and Test Modes on page
details.
www.masimo.com
Chapter 4: Operation
, PI and PR only. Using this mode
2
4341
for details.
39
2
4139
for details.
4139
for