Masimo Pronto-7 Operator's Manual page 39

Spot chek pulse co-oximeter
Hide thumbs Also See for Pronto-7:
Table of Contents

Advertisement

Pronto-7
Viewing Select Parameters
Upon successful completion of a spot check test, the default parameters
shown on the firs page of the Test Results screen are SpHb, SpO
PI.
If the user would like the Test Results screen to display specific parameters,
on page 2, perform the following:
1.
Tap the Main Menu icon.
2.
Tap the Display icon.
3.
Toggle parameters accordingly to display the parameters preferred
on page 1. Parameters listed under Display Parameters in green are
those that may be displayed on the Test Results screen page 1,
those in gray will be displayed on page 2.
4.
Tap the green check mark icon to accept changes.
Information about each parameter is available. Tap directly on the parameter
for access. A pop-up window appears for the selected parameter.
Viewing SpHct
There are two ways to access SpHct.
To access prior to conducting a test, perform the following:
1.
Tap the Main Menu icon.
2.
Tap the Display icon.
3.
Toggle to SpHct under Units of Measure.
4.
Tap the green check mark icon to accept changes.
To access upon completion of a test, perform the following:
1.
Tap the SpHb text next to the SpHb numeric result.
2.
The text will spin to show the resulting SpHct measurement. Tap on
the SpHct text next to the SpHct numeric result to return to SpHb.