Gas Supplies - Agilent Technologies 240 Series AA User Manual

Including zeeman
Table of Contents

Advertisement

Introduction
Non-Zeeman
240 series AA
T2.5 A H250 V, IEC 127 sheet 5, 5 x 20 mm (100–120 & 220–240 VAC)
280FS AA
T4 A H250 V, IEC 127 sheet 5, 5 x 20 mm (100–120 & 220–240 VAC)
Zeeman
240Z/280Z AA
15 A long delayed-action circuit breaker with a thermal cutout.
T3.15 A H 250 V, IEC 127 sheet 5, 5 x 20 mm (208–240 VAC)
T3.15 mA L 250 V, IEC 127 sheet 2, 5 x 20 mm (208–240 VAC)
T1 A L 250 V, IEC 127 sheet 2, 5 x 20 mm (208–240 VAC)
T6-3 AL 250 V, IEC 60127-2 sheet 3, 5 x 20
T800 mA L 250 V, IEC 127 sheet 2, 5 x 20 mm (208–240 VAC)
For safety reasons, any other internal fuse or circuit breaker is not operator
NOTE
accessible and should be replaced only by Agilent field service engineers.
Fuse information on the rear of the instrument is the most up to date.

Gas supplies

C
H
2
Instrument grade
>99.5% pure
Allowed range
65–100 kPa
(9.5–14.5psi)
Recommended
75 kPa (11 psi)
Typical flow rate
0–10 (L/min)
Connection
1.8 m (6 ft) length
of 6.3 mm (1/4 in)
ID red hose with
9/16 in-18UNF L/H
thread fitting#
Adaptors are available
#
Other gas connections
Sample compartment:
36
Air
2
Must be clean and
dry. (Air filter to be
used if required)
245–455 kPa
(35–65 psi)
350 kPa (50 psi)
13.5-20 (L/min)
1.8 m (6 ft) length
of 6.3 mm (1/4 in)
ID black hose with
9/16 in-18UNF
fitting#
Push-on Air/N
Push-on C
N
O
2
Instrument grade
>99.5% pure
245–455 kPa
(35–65 psi)
350 kPa (50 psi)
11–16 (L/min)
1.8 m (6 ft) length
of 6.3 mm (1/4 in)
ID blue hose with
3/4 in-16UNF
fitting#
O connector for burner
2
H
connector for burner
2
2
Agilent 240/280 Series AA (including Zeeman)
Air purge
245–455 kPa
(35–65 psi)
10 (L/min)
Barb to take
6.3 mm (1/4 in)
hose