Troubleshooting
9.2
Warning reference
Warning
Cause
number
#101
Datalogger in Stack mode is 100 % full and is
not logging any more data.
#102
Datalogger in Stack mode is 90 % full.
#103
Data error in the device. Functions such as
#106
the datalogger or dongle functions may no
longer be available.
#104
No valid date or time set.
#107
Battery almost used up.
#110
SW dongle has been removed without
#111
logging out.
#112
#113
Alarms disabled.
#164
Sensor warm-up phase has not ended yet.
#165
Increased measurement error must be
expected.
#167
Calibration interval expired.
#170
28
Remedy
Read out the data, clear the datalogger and start again.
Read out the data as soon as possible, clear the datalogger and
start again.
Reset the device to factory settings (see chapter 8.2.6 on page 21).
If this fault occurs again:
Have the device checked by DrägerService.
Set time and date (see chapter 8.2.3 on page 20).
Have the device checked by DrägerService.
Disable SW dongle (see chapter 8.2.7 on page 21).
Enable alarms (see chapter 8.2.1 on page 19).
Wait until the sensor has warmed up.
Recalibrate the device (see chapter 5.1 on page 15).
Dräger Polytron 8000