3.

CONNECTION

The supply lead for feeding the instrument should not be in the proximity of low-potential signals.
Contactors, motors with larger input and other efficient elements should not be in the proximity of the instrument.
The lead into the instrument input (the measured quantity) should be in sufficient distance from all power leads and appliances.
Provided this cannot be secured, it is necessary to use shielded leads with connection to ground.
The instruments are tested in compliance with standards for use in industrial area, yet, we recommend to abide by the above
mentioned principles.
OM 36AC
GND
MEASURING RANGE
Type
Input 1
OM 36 DC - U
±199,9 mV; ±1,999 V; ±19,99 V
±199,9 νA ±1,999 mA; ±19,99 mA
OM 36 DC - I
±199,9 mA, ±1,999 A; ±5,00 A
OM 36 AC - U
0...199,9 mV; 0...1,999 V; 0...19,99 V
0...1,999 mA; 0...19,99 mA; 0...199,9 mA
OM 36 AC - I
0...1,999 A; 0...5,00 A
1
2
3
4
-
-
OM 36PM
INPUT 2
INPUT 1
INPUT
Input 2
±199,9 V; ±300 V
0...199,9 V; 0...300 V
5
6
7
8
9
10 11
-
-
OM 36OHM
3. CONNECTION
OM 36DC
OM 36RTD
5