If the display speed is slower after average sampling time is enabled - this is normal
if average sampling is carried out more than 32 times. You can reduce the intervals
of average sampling.
Waveform display in ladder shape - this is normal. The reason may be the
horizontal time base range is too low. If the display type is vector, the connection
between sampling dots may produce a ladder waveform. Set the display type to dot
to solve this issue.

SPECIFICATION

Sampling modes
Sampling rates
Average value
Input coupling
Input impedance
Probe attenuation
Maximum input voltage
Time delay between channels
(typical)
Waveform interpolation
Record length
Scanning range (s/div)
Accuracy of sampling rate and
delay time
Time interval (rT)
measurement accuracy
(full bandwidth)
Sampling
Real time
72-2650
72-8710A
72-8705A
72-8225A
72-10510
When all channels have made N samplings simultaneously,
N is selectable from 2, 4, 8, 16, 32, 64, 128 and 256
Input
72-2650
72-10510 72-8225A
72- 8705A 72-8710A
400V(DC+AC peak, 1MΩ input impedance)
Horizontal
2 x 512k sampling point
72-8705A 72-8710A
72-8225A
72-10510
±50ppm (any time interval ≥1m)
Single: ±(1 sampling time interval + 50ppm X reading + 0.6ns)
>16 average values: ±(1 sampling time interval + 50ppm x
24
Real time
1GS/s
1GS/s
1GS/s
500MS/s
250MS/s
DC, AC, GND
1±2%MΩ with 21pF±3pF
1±2%MΩ with 24pF±3pF
1X, 10X, 100X, 1000X
150ps
Sin (x) / x
reading + 0.4ns)
Equivalent
50GS/s
50GS/s
50GS/s
25GS/s
----
2ns/div-50s/div
5ns/div-50s/div
10ns/div-50s/div