Problem
The controller
• Damage has occurred to an
does not show
attached measurement
connected
device or the device cable.
sensors.
• There is a communication
failure between the controller
and one or more of the
attached measurement
devices.
The controller
The sensor settings are not
does not show
correct.
connected
sensors.

Accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Accessories
Description
Sunroof for controller
Sunroof with mounting hardware
Sunroof adapter mounting plate for CD500 and CD300 controllers
Possible cause
W A R N I N G
• Make sure that the measurement device is connected
to the controller.
• Examine the measurement device and the device cable
for damage. Make sure that the measurement device is
operational. If damage is found, contact technical
support
• Set the controller to off and then to on again.
Make sure that the sensor settings on the sensor menu
are correct. Refer to
Connect a sensor
If connected sensors are not Modbus, do the steps that
follow:
1. On the main menu select Sensor.
2. Select Protocol>Trios.
Solution
on page 17.
Item no.
LZX958
LZX957
LXZ529.99.00027