Calibration; Calibrate The Vertical Scale - Promax OD-590 User Manual

Digital oscilloscope
Hide thumbs Also See for OD-590:
Table of Contents

Advertisement

Available languages

Available languages

USER'S MANUAL. OD-590/91/92

8. Calibration

8.1 Calibrate the vertical scale

* Run Calibration under two conditions
1.
When using Oscilloscope in a new environment, such as field
measurement.
2.
When the temperature changes more than 5°C.
Operation
Utility
More
Make sure the environment fits these conditions.
1.
Connect the rear panel Calibration output to Channel1. (BNC male —
male connector)
Calibration Output
CAL
2.
Press the Utility key→F5→F1→F1
3.
Press F5 and start the calibration. It takes approximately 2 minutes.
4.
When completed, switch the connection to channel 2. Repeat the
above process for the whole channel.
Page 103
Self Cal
F 5
Menu
Temperature: 26 ± 5°C
Relative humidity: ≤ 80%
Connect to channel 1
2 V
G D S-2 1 0 4
D i gi t al S t o r a g e Osc i l l osc op e
1 00 MH z 1G S a / s
F 1
Vertical
HORIZONTAL
Ut i lity
P OS I T I ON
H OR I
ME N U
ON /
STBY
T I ME / D I V
TRIGG ER
V A R I A B L E
He lp
L E V E L
M E N U
VERT ICA L
P OS I T I ON
P OS I T I ON
P OS I T I ON
P OS I T I ON
C H 1
M AT H
C H 2
C H 3
C H 4
V OL T S / D I V
V OL T S / D I V
V OL T S / D I V
V OL T S / D I V
CH1
CH2
CH3
CH4
1
MΩ
16pF
1
MΩ
16pF
300V
CA T
300V
CA T
MA X. 3 00Vpk
MA X. 3 00Vpk
F 1
01-2007

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Od-591Od-592

Table of Contents