4. Startup
Checking the voltage
21
26
27
14
16
20
(continued)
22
25
23
24
Fig. 12
17
Fig. 13
The voltage selector is located above the main
power switch, on the left-rear side of the instru-
ment
(Fig.
7). The voltage setting is displayed in
the viewing window (22).
• Insert a small screwdriver into the cutout
(25) and carefully pry out the insert.
• Remove the voltage selector housing (21)
together with the fuses (23). Remove the
voltage selector block (24) (white) and insert
it again such that the correct local voltage is
21
displayed in the viewing window (22).
• Reinsert the voltage selector housing with
the block and fuses and push it in until it
engages (audible click).
Connecting the control panel
The connecting cable (15) of the control panel is
fixed to the microtome. It may not be discon-
nected.
• Insert the plug (14) of the connecting cable
(15) into the connecting jack (16) on the rear
of the control panel.
• To fasten the plug, tighten the two screws
15
(17).
Instruction manual V 1.6 – 10/2009