Part 2: adjustment and operation measured variable oxygen
Table of Contents

Advertisement

Complete Operating Menu / Description
Standard signal input
lower signal limit
Allocated measuring
range
Checkout time
Measurement parameters
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
meas. parameter
atmospheric air
pressure
=
1013 mbar
Water temperature
Air temperature
at calibration
Atmospheric air pressure
Height above sea level
Relative humidity
Salt content of water
To retain the prevailing O
of water" menu option (arrow keys): the DUCLOMETER
O
-saturation.
2
Possible Values
Initial value
Increment
4 mA
0 mA
4 mA
0...10.00 ppm
0.01 ppm
0...20.00 ppm
0.01 ppm
0...50.00 ppm
0.01 ppm
0...100 %
0.1 %
0...200 %
0.1 %
0...500 %
0.1 %
off
1 s
meas. parameter
temp. of water
=
10.0 °C
O
:
100 %
2
meas. parameter
height about
sea level
=
300 m
Possible Values
Initial value
Increment
10.0 °C
0.1 °C
20.0 °C
0.1 °C
1013 mbar
1 mbar
300 m
1 m
100 %
1 %
0 g/l
1 g/l
-saturation, enter the actual water temperature in the "meas. parameter, temp.
2
Lower value
Upper value
-1.00 ppm
11.00 ppm
-2.00 ppm
22.00 ppm
-5.00 ppm
55.00 ppm
-10.0 %
110.0 %
-20.0 %
220.0 %
-50.0 %
550.0 %
1 s
9999 s
meas. parameter
temp. of air
at calibration
=
20.0 °C
meas. parameter
relative humidity
=
100 %
Lower value
Upper value
0.0 °C
40.0 °C
0.0 °C
0.0 °C
300 mbar
1180 mbar
0 m
9999 m
0 %
100 %
0 g/l
50 g/l
®
D1C immediately calculates the relative
Remarks
Constant measurement signal
results in message and alarm.
Function off = 0 s
meas. parameter
salinity of
water
=
0 g/l
D1C2-O2-032-GB
Remarks
15