Easidew Online User's Manual
4.3
Fault Conditions
Message
Displayed
ErrL
Sbr
ErrH
outR
rurC
Michell Instruments
Cause
Sensor failure
Instrument failure
Sensor failure or break in
sensor connection
Gas is wetter than +20°Cdp Check gas source supply
Sensor contaminated
Input out of range
Reverse input connection
MAINTENANCE
Action
Check power supply to transmitter. Check
transmitter cable for continuity/damage.
Rectify/replace cable
Refer to Michell or local representative for
repair
Check transmitter cable for continuity/
damage.
Rectify/replace cable
Replace/re-calibrate transmitter
Check gas source supply.
Re-calibrate/replace transmitter
Swap
input
connections
transmitter to the monitor
from
the
43