OPERATION
3.4
Operating Temperature / ppm
3.4.1

Temperature Range Default

The default temperature unit for the Easidew Online instrument is in degrees Celsius.
This is indicated by the °Cdp LED indicator. The default settings associated with this
temperature scale are as follows:
Span -100 to +20°Cdp
Lower and upper span limits -100 and +20 (display fl ashes outside this
range)
Minimum alarm set-point -100°Cdp
Maximum alarm set-point +20°Cdp
To change the range to Fahrenheit follow the procedures in Sections 3.4.2 and 3.4.3.
32
Range
V
To range the instrument for °F, all the above parameters need
to be changed to their Fahrenheit equivalent values (-148 and
+68°F). It is not suffi cient just to change the °F/°C units.
Easidew Online User's Manual