To Replace the Power Line Fuse

The power line fuse is located within the multimeter's fuse- holder assembly
on the rear panel. The multimeter is shipped from the factory with a
power- line fuse installed (according to country of destination). Refer to
Table 5-
same size and rating.
Table 5-1 Type of supplied fuse (according to country of destination)
1 Disconnect the power cord. Depress tab 1 and
3 Rotate the line voltage selector and reinstall
U3402A User's and Service Guide
1. If you determine that the fuse is faulty, replace it with one of the
Type of fuse (time-lag, low breaking fuse)
0.25A, 250V, 5x20 mm
0.125A, 250V, 5x20 mm
2 and pull the fuse holder from rear panel.
it back, so that the correct voltage appears in
the fuse holder window.
100, 120, 220 (230), or 240 Vac
Disassembly and Repair
Input line voltage
100 V to 120 V
220 V to 240 V
2 Remove the line voltage selector from fuse
holder assembly.
2
1
4 Replace fuse holder assembly in rear panel.
5
Part number
A02-62-25592-3U
A02-62-25648-1U
79