5 .11 Start-up procedure

5 .11 .1
Level measurement
The filling level is calculated by the device due to the distance of the filling material surface to the
sensor.
Sub menu Basic Setting
Commissioning
Min/Max Adjustment
Sub menu Display
Indicator value fill level
5 .11 .1 .1
Example
Measurement task
Distance Sensor - container bottom:1,8m
Minimum filling level: 0m >> distance 1,8m – 0m = 1,8m
Maximum filling level:1,4m >> distance 1,8m – 1,4m = 0,4m
Settings
Process commissioning
Min/Max Adjustment: 1,8m = 0% / 0,4m = 100%
A - Filling level
B - Distance sensor – filling material surface
C - Distance sensor – container bottom