Cleaning The Pressure Transducer - vacuubrand ME 4C NT VARIO Instructions For Use Manual

Chemistry diaphragm pumps. chemistry vacuum system. chemistry pumping units
Hide thumbs Also See for ME 4C NT VARIO:
Table of Contents

Advertisement

Cleaning the pressure transducer

Attention: Never use a pointed or sharp-edged tool to
NOTICE
clean the pressure transducer.
Never touch the ceramic diaphragm of the pressure
transducer with hard objects.
- Fill the measurement chamber with a solvent (e.g.,
- Drain the solvent and dispose of in accordance with
- Rinse the measurement chamber several times with al-
- Allow the pressure transducer to dry.
- Readjust the pressure transducer if necessary.
Readjustment of the controller CVC 3000
See section "Readjustment of CVC 3000", pg. 78.
page 81 of 129
benzene) and allow sufficient cleaning time. Observe
all regulations concerning usage and disposal of sol-
vents!
regulations. Repeat cleaning if necessary.
cohol in order to remove all solvent residues.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents