During operation

➨ Vent and dispose of potentially dangerous gases or
+ Due to the high compression ratio, the pump might gen-
+ Maximum ambient temperature: 104 °F (40 °C)
• If the pump is installed at an altitude of more than 6500
• Check compatibility with the maximally permitted
Do not start the pump if the pressure difference between
NOTICE
inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).
Attempts to start the pump at higher pressure difference
may cause stalling and damage of the motor.
If pumping condensable vapors (water vapor, solvents,
etc.), let the pump run with gas ballast to help purge any
condensation in the pump.
Prevent internal condensation, transfer of liquids or dust.
The diaphragms and valves will be damaged, if liquids are
pumped in significant amounts.
Check the pump regularly for external soiling and depos-
page 44 of 129
vapors at the outlet of the pump appropriately.
erate overpressure at the outlet. Check pressure com-
patibility with system components (e.g., exhaust tubing
or exhaust valve) at the outlet. Ensure that the pump
outlet is neither blocked nor restricted.
Check the maximum temperatures, if installing the
pump in a cabinet or a housing. Make sure ventilation
is adequate, especially if the ambient temperature is
elevated.
ft (2000 m) above mean sea level, check compatibility
with applicable safety requirements, and adopt suitable
measures. There is a risk of the motor overheating due
to insufficient cooling.
pressure at outlet and the maximum pressure differ-
ence between inlet and outlet ports.